| ECOLINE VC-280 / 281 |
 |
|
| ..VC-280 for tubing with 1.6 mm wall thickness |
| ..VC-281 for tubing with 2.4 mm wall thickness |

|
 |
| Tubing type for VC-280 |
Standard tubing for 1.6 mm wall thickness
1.6 11.1 mm (1/16 7/16") inner diameter |
| Tubing type for VC-281 |
Standard tubing for 2.4 mm wall thickness 4.8 9.5 mm (3/16 3/8 ") inner diameter |
| Tube-bed |
Self-centring tube-track |
| Rotor speed |
3.5 350 rpm, adjustable in 1% steps |
| Speed setting |
1-99%, resolution 1%, 2-digit potentiometer |
| Speed control |
Closed loop control for pressure independent speed |
| Analog input |
Speed control (05 or 010V / 020 or 420mA) |
| Digital input (TTL level) |
Start/Stop, rotation direction |
| Motor type |
DC motor |
| Power consumption |
100 W |
| Power supply |
230VAC/50Hz, 115VAC/60Hz, selectable |
| Differential pressure |
- 1.5 bar (21.7 psi) possible (depending on tubing material);
- small tubing diameters may allow higher pressures |
| Protection rating |
IP 30 |
| Depth / Width / Height |
256 x 169 x 138 mm (10 1/8" x 6 3/4" x 5 3/8") |
| Weight |
5.2 kg (12 lbs) |
| Accessories |
- foot switch
- interchangeable rotors |
|
|
Overview
ECOLINE pumps |
ISMATEC SA, Labortechnik - Analytik, Feldeggstrasse 6, 8152 Glattbrugg, Switzerland,
www.ismatec.com, sales.ismatec@idexcorp.com, Inquiry form, Phone and fax numbers
|