ECOLINE VC-280 / 281
Ecoline VC-280
..VC-280 for tubing with 1.6 mm wall thickness
Order No. Channels

Flow rates
ml/ min per channel

Pump
rollers
ISM 1078 1 1.7 – 5'400 2

..VC-281 for tubing with 2.4 mm wall thickness
Order No. Channels

Flow rates
ml/ min per channel

Pump
rollers
ISM 1085 1 13 – 4'600 2

> Operating Manual (PDF) > Interchangeable Rotors
Tubing type for VC-280 Standard tubing for 1.6 mm wall thickness
1.6 11.1 mm (1/16 7/16") inner diameter
Tubing type for VC-281 Standard tubing for 2.4 mm wall thickness
4.8 9.5 mm (3/16 3/8 ") inner diameter
Tube-bed Self-centring tube-track
Rotor speed 3.5 350 rpm, adjustable in 1% steps
Speed setting

1-99%, resolution 1%, 2-digit potentiometer

Speed control Closed loop control for pressure independent speed
Analog input Speed control (05 or 010V / 020 or 420mA)
Digital input (TTL level) Start/Stop, rotation direction
Motor type DC motor
Power consumption 100 W
Power supply 230VAC/50Hz, 115VAC/60Hz, selectable
Differential pressure - 1.5 bar (21.7 psi) possible (depending on tubing material);
- small tubing diameters may allow higher pressures
Protection rating IP 30
Depth / Width / Height

256 x 169 x 138 mm (10 1/8" x 6 3/4" x 5 3/8")

Weight 5.2 kg (12 lbs)
Accessories - foot switch
- interchangeable rotors

Overview ECOLINE pumps


ISMATEC SA, Labortechnik - Analytik, Feldeggstrasse 6, 8152 Glattbrugg, Switzerland,