   Technology
   What's New

 

ECOLINE VC-360
Select the tubing i.d. according to the required flow rate


ecovc360_mit.jpg (14033 Byte)
ECOLINE VC-360
ISM 1076

Flow rates
ml/min

min. max.
0.25 25
0.9 90
3.5 350
7.7 770
13 1300

Standard tubing
1.6 mm wall thickness

i.d. mm

0.8
1.6
3.2
4.8
6.4

Above mentioned flow rates are approximate values, determined with water, at 22° C (72° F), without differential pressure, with Tygon® tubing.