| CA
pump-heads
Select the tubing i.d. according to the required flow rate
|
|

CA 4 4 channels ISM 721

CA 8 8 channels ISM 732

CA 12 12 channels ISM 733
|
Flow
rates
ml/min per channel |
| 0.002 |
0.31 |
| 0.004 | 0.94 |
| 0.008 |
1.8 |
| 0.019 | 4.5 |
| 0.025 |
6.1 |
| 0.034 | 8.2 |
| 0.042 |
10 |
| 0.053 | 13 |
| 0.074 |
18 |
| 0.10 | 24 |
| 0.11 |
27 |
| 0.13 | 31 |
| 0.14 |
35 |
| 0.16 | 38 |
| 0.18 |
42 |
| 0.20 | 47 |
| 0.23 |
55 |
| 0.26 | 62 |
| 0.30 |
71 |
| 0.33 | 78 |
| 0.36 |
86 |
| 0.43 | 100 |
| 0.51 |
120 |
| 0.62 | 150 |
| 0.74 |
180 |
| 0.94 | 230 |
|
| 0.13 |
| 0.19 |
| 0.25 |
| 0.38 |
| 0.44 |
| 0.51 |
| 0.57 |
| 0.64 |
| 0.76 |
| 0.89 |
| 0.95 |
| 1.02 |
| 1.09 |
| 1.14 |
| 1.22 |
| 1.30 |
| 1.42 |
| 1.52 |
| 1.65 |
| 1.75 |
| 1.85 |
| 2.06 |
| 2.29 |
| 2.54 |
| 2.79 |
| 3.17 |
|
| For
BVP-Standard drive, the min. flow rate values must be multiplied by factor 2.4 |
| Above
mentioned flow rates are approximate values, determined with water, at 22° C (72° F), without differential pressure, with Tygon® tubing. |
| | |
ISMATEC SA, Labortechnik - Analytik, Feldeggstrasse 6, 8152 Glattbrugg, Switzerland,
www.ismatec.com, sales.ismatec@idexcorp.com, Inquiry form, Phone and fax numbers
|