| Pump-head SB 2V & SB 3V
Select the tubing i.d. according to the required flow rate |

SB 2V ISM 734 plus ISM 010 |
Flow
rates
ml/min per channel |
| 1.1 | 260 |
| 2.3 | 550 |
| 3.7 | 890 |
| 4.6 | 1100 |
| |
|
|
| 3.2 | 1/8 |
| 4.8 | 3/16 |
| 6.4 | 1/4 |
| 8.0 | 5/16 |
| |
|
|
| For
BVP-Standard drive, the min. flow rate values must be multiplied by factor 2.4 |

SB 3V ISM 734 plus ISM 011 |
| Flow
rates ml/min per channel |
| 0.09 |
22 |
| 0.26 |
63 |
| 0.99 |
240 |
| 2.2 |
530 |
| |
|
|
| 0.8 |
1/32 |
| 1.6 |
1/16 |
| 3.2 |
1/8 |
| 4.8 |
3/16 |
| |
|
|
| For
BVP-Standard drive, the min. flow rate values must be multiplied by factor 2.4 |
| Above
mentioned flow rates are approximate values, determined with water, at 22° C (72° F), without differential pressure, with Tygon® tubing. |
|
| |
ISMATEC SA, Labortechnik - Analytik, Feldeggstrasse 6, 8152 Glattbrugg, Switzerland,
www.ismatec.com, sales.ismatec@idexcorp.com, Inquiry form, Phone and fax numbers
|